Fix missing anchor3
Build Sphinx Documentation / Build Documentation from PR Branch (pull_request) Successful in 2m14s

This commit is contained in:
2025-10-31 17:49:09 +08:00
parent f968c8ff85
commit 9d53cb7e60
@@ -40,7 +40,7 @@
| MeasureV | FVMV/FIMV @ 10Ω All Channel| | [Note2](#custom-anchor2) | | MeasureV | FVMV/FIMV @ 10Ω All Channel| | [Note2](#custom-anchor2) |
| MeasureI | FVMI @ 10Ω All Channel| | [Note2](#custom-anchor2) | | MeasureI | FVMI @ 10Ω All Channel| | [Note2](#custom-anchor2) |
|ClampV| FIMV @ 10Ω All Channel|| Pass! | |ClampV| FIMV @ 10Ω All Channel|| Pass! |
|ClampI| FVMV/FVMI @ 10Ω All Channel| | Fail! [Note3] | |ClampI| FVMV/FVMI @ 10Ω All Channel| | Fail! [Note3](#custom-anchor3) |
## 3. 问题 ## 3. 问题
@@ -55,7 +55,7 @@
4. 通过ATE_Debugger发送指令:newdpsfunc("newdpsinitvalue","disable"); 4. 通过ATE_Debugger发送指令:newdpsfunc("newdpsinitvalue","disable");
5. 执行测量 5. 执行测量
3. 在ClampI模式下(FVMV/FVMI),若Clamp值与Force值接近(如80mA档位Clamp60mA,Force70mA;2mA档位Clamp1mA,Force1.5mA)则Force输出与Sense管脚及对应信号链均会出现震荡,且在极端情况下震荡 ***可能不收敛*** 3. 在ClampI模式下(FVMV/FVMI),若Clamp值与Force值接近(如80mA档位Clamp60mA,Force70mA;2mA档位Clamp1mA,Force1.5mA)则Force输出与Sense管脚及对应信号链均会出现震荡,且在极端情况下震荡 ***可能不收敛*** <a id="custom-anchor3"></a>
![ClampI](../../_images/NewDPS/clamp_003.png) ![ClampI](../../_images/NewDPS/clamp_003.png)