From 9d53cb7e603c754231caa27784fe2855d80b9a37 Mon Sep 17 00:00:00 2001 From: YuYang Shen Date: Fri, 31 Oct 2025 17:49:09 +0800 Subject: [PATCH] Fix missing anchor3 --- source/Lab_Report/Custom_Board/NewDPS_Verify.md | 4 ++-- 1 file changed, 2 insertions(+), 2 deletions(-) diff --git a/source/Lab_Report/Custom_Board/NewDPS_Verify.md b/source/Lab_Report/Custom_Board/NewDPS_Verify.md index 941111a..e322622 100644 --- a/source/Lab_Report/Custom_Board/NewDPS_Verify.md +++ b/source/Lab_Report/Custom_Board/NewDPS_Verify.md @@ -40,7 +40,7 @@ | MeasureV | FVMV/FIMV @ 10Ω All Channel| | [Note2](#custom-anchor2) | | MeasureI | FVMI @ 10Ω All Channel| | [Note2](#custom-anchor2) | |ClampV| FIMV @ 10Ω All Channel|| Pass! | -|ClampI| FVMV/FVMI @ 10Ω All Channel| | Fail! [Note3] | +|ClampI| FVMV/FVMI @ 10Ω All Channel| | Fail! [Note3](#custom-anchor3) | ## 3. 问题 @@ -55,7 +55,7 @@ 4. 通过ATE_Debugger发送指令:newdpsfunc("newdpsinitvalue","disable"); 5. 执行测量 -3. 在ClampI模式下(FVMV/FVMI),若Clamp值与Force值接近(如80mA档位Clamp60mA,Force70mA;2mA档位Clamp1mA,Force1.5mA)则Force输出与Sense管脚及对应信号链均会出现震荡,且在极端情况下震荡 ***可能不收敛*** +3. 在ClampI模式下(FVMV/FVMI),若Clamp值与Force值接近(如80mA档位Clamp60mA,Force70mA;2mA档位Clamp1mA,Force1.5mA)则Force输出与Sense管脚及对应信号链均会出现震荡,且在极端情况下震荡 ***可能不收敛*** ![ClampI](../../_images/NewDPS/clamp_003.png)